Anesthesia machine parts, breathing circuits and the leak test
Oxygen leaves the cylinder or pipeline, drops to a working pressure at the regulator, is metered by the flowmeter, picks up inhalant in the vaporizer and exits at the common gas outlet into the breathing circuit. Rebreathing (circle) circuits reuse gas and need CO2 absorbent and one-way valves. Non-rebreathing circuits suit small patients. Leak test before every patient.
Parts of the anesthesia machine, in gas-path order
Every anesthesia machine does the same job: it takes compressed gas at high pressure, lowers that pressure, and mixes the gas precisely with a potent inhalant so it can reach the patient through a breathing circuit. One teaching handout splits the machine into four parts: a high pressure system where the gas supply attaches, a low pressure system where oxygen and inhalant are mixed, the breathing system, and the scavenging system.[3]
To learn the parts, follow the oxygen from its source through the machine and circuit to the patient, then to the bag and out through the scavenger.[1]
Cylinders (oxygen cylinders are green in the US) or a pipeline. Cylinders attach by a pin-indexed yoke so the wrong gas cannot be connected.
Pressure gauge
Shows the pressure in the cylinder or pipeline. For oxygen, pressure falls steadily as the cylinder empties.
Pressure regulator
Reduces cylinder pressure to a working level of about 45 psi so high pressure gas never reaches the flowmeter, and keeps flow steady as the cylinder empties.
Flowmeter
Controls and shows the gas flow rate. A bobbin or ball rises in a tapered tube; the higher the flow, the higher it rises.
Vaporizer
Sits downstream of the flowmeter, holds liquid inhalant and adds a set concentration of vapor to the gas. Precision vaporizers are agent specific.
Common gas outlet
Where the mixed oxygen and inhalant leave the machine to enter the breathing circuit. A frequent source of leaks.
Oxygen flush valve
Sends pure oxygen straight to the common gas outlet, bypassing the flowmeter and vaporizer.
Oxygen supply, regulator and flowmeter: what to check
Before the day starts, confirm the oxygen tank or hospital supply reads more than 200 psi, or that the oxygen generator is working. A full oxygen cylinder is filled to about 2,200 psi.[1],[3]
Most machines have a pressure regulator (pressure reducing valve) built in. When both a pipeline and a cylinder are attached, the machine draws from the pipeline first, because pipeline gas arrives at about 50 psi and the regulated cylinder gas at about 45 psi.[3]
Flowmeters are calibrated for one gas each. Read the flow at the top of a bobbin, or at the middle of a ball. The control knob is delicate, so turn it gently and never force it closed.[3],[5]
The vaporizer and the oxygen flush valve
A precision vaporizer splits incoming gas into two streams. One passes through a chamber above the liquid inhalant and picks up vapor; the other bypasses it. The dial sets the ratio between the two, which sets the concentration delivered. It sits out of the circuit, on the machine, and is filled with liquid inhalant as part of daily setup.[1],[3],[5]
The oxygen flush delivers pure oxygen, without inhalant, at 35 to 60 or up to 75 L/min, depending on the source. That volume can over-pressurize small lungs. It should never be used in a patient breathing through a non-rebreathing circuit, because it sends a high pressure flow straight into the airway.[1],[3],[5]
In a rebreathing circuit the patient breathes some of the same gas again, and CO2 is removed as the gas passes through absorbent such as soda lime. A circle circuit has a CO2 absorbent canister, a Y-piece, inspiratory and expiratory breathing tubes, two one-way valves, a fresh gas inlet, a pressure manometer, a pop-off valve and a reservoir bag.[4]
The one-way (unidirectional) valves send gas toward the patient on inspiration and away on expiration, so exhaled gas cannot be breathed again before it passes through the absorbent. A faulty valve lets CO2 build up. You can check them by breathing through the circuit and watching the correct valve move on each breath.[1],[4]
Swap in fresh absorbent once a canister has seen 8 hours of use. Exhausted soda lime granules turn hard and do not crush. Do not rely on the color indicator alone: it can change color too late and can return to its original color when the canister is not in use.[1],[2],[4]
The reservoir bag holds gas for the next breath, shows you the breathing pattern, and lets you give breaths by hand. Small animal bags commonly range from 0.5 to 5 liters; one handout gives 500 ml for cats and small dogs and 2 liters for a 20 kg dog.[4]
The manometer shows circuit pressure in cm H2O. The adjustable pressure-limiting (APL) valve, also called the pop-off valve, lets waste and excess gas leave the circuit toward the scavenger. Left closed, it lets pressure build in the patient's lungs. AAHA names a manometer and a separate safety pop-off valve as essential on every machine.[1],[2],[4]
Rebreathing vs non-rebreathing circuits: how to choose
Non-rebreathing circuits, such as the Bain and the Ayre's T-piece, have no CO2 absorbent. The patient breathes from the reservoir and exhaled gas is not reused; CO2 is cleared by a high fresh gas flow and low dead space. They offer less resistance to breathing and low equipment dead space, which matters in small patients.[1],[2],[4]
AAHA guidance describes non-rebreathing circuits for cats and small dogs under 3 to 5 kg, and circle circuits for patients over 3 to 5 kg. A pediatric rebreathing circuit can be used for smaller patients if a non-rebreathing circuit is not available. Feline guidance names a non-rebreathing circuit for very small cats under 3 kg. One teaching handout puts the cutoff near 6 kg, so follow your hospital's protocol.[1],[2],[4]
Rebreathing and non-rebreathing circuits compared[1],[4]
Feature
Rebreathing (circle)
Non-rebreathing (Bain, T-piece)
Usual patient size (AAHA)
Over 3 to 5 kg
Under 3 to 5 kg
How CO2 is removed
CO2 absorbent
High fresh gas flow
One-way valves
Yes
No
Resistance to breathing
Higher
Lower
Heat and moisture
Conserved
Cold, dry gas; more heat loss
Inhalant use and pollution
Lower
Higher
How to do an anesthesia machine leak test
Leak test the machine and circuit before every patient and after changing any circuit part. It confirms oxygen will reach the patient and that little inhalant will escape.[1],[2]
Close the APL (pop-off) valve. Block the patient end of the circuit. Raise the oxygen flow until the manometer reads 20 to 30 cm H2O, then turn the oxygen off. The needle should hold steady. If it does, open the pop-off valve first, then release the patient end. If the pressure falls, take the machine out of use and troubleshoot it.[1]
Scavenging: active vs passive
The scavenger collects waste gas leaving the APL valve so staff are not exposed. Active systems use a fan or vacuum to draw gas outside and need an air break so suction is not applied to the patient's breathing circuit. Passive systems rely on the patient's own breath to push gas out through tubing or into an activated charcoal canister.[4],[5]
Active scavenging is far more effective than passive. Weigh charcoal canisters regularly and replace them when they reach a set weight. They do not absorb nitrous oxide.[1],[4],[5]
Worked examples
Example 1: A falling needle during the morning pressure check
You are setting up a circle circuit for a 27 kg Labrador's dental. You close the pop-off valve, thumb over the Y-piece, and fill the circuit to 25 cm H2O. After you turn off the oxygen, the needle drifts to 12 cm H2O in about 20 seconds.
1Step 1: Know the pass mark: once the oxygen is off, the manometer reading should stay steady.[1]
2Step 2: Compare: a drop from 25 to 12 cm H2O means gas is escaping somewhere in the machine or circuit.[1]
3Step 3: Act: do not anesthetize this patient on this machine. Pull it from use and troubleshoot, starting with likely leak points such as the common gas outlet connection.[1],[3]
Answer
The machine fails the leak test. Leaks reduce oxygen delivery and expose staff to inhalant, so the machine stays out of use until the leak is found and fixed.[1],[3]
Example 2: Two patients, two circuits
Your list has a 2.2 kg Siamese cat for a mass removal and a 31 kg Boxer for a cruciate repair. Your hospital stocks Bain circuits and adult circle circuits.
1Step 1: Cat: at 2.2 kg it is a very small cat, under the 3 kg line in feline guidance and under AAHA's 3 to 5 kg range, so use the Bain (non-rebreathing) circuit.[1],[2]
2Step 2: Why: a non-rebreathing circuit has lower resistance and less dead space, which matters for a small patient.[1]
3Step 3: Dog: at 31 kg it is well over 3 to 5 kg, so use the circle (rebreathing) circuit.[1]
4Step 4: For the cat, remember the oxygen flush is off limits while it is connected to the Bain.[1],[2]
Answer
Bain for the cat, circle for the dog. Weight drives the choice because resistance and dead space matter most in small patients.[1],[2]
Example 3: Absorbent that still looks fresh
The canister log for the surgery machine shows 9 hours of use since the last change. The granules look their original color, but when you press a few they are hard and will not crush.
1Step 1: Check the time: a canister is due for fresh absorbent at 8 hours, and this one has run 9.[1],[2]
2Step 2: Check the texture: hard, non-crushable granules mean the soda lime is exhausted.[4]
3Step 3: Weigh the color: indicators can change too late and can revert to the original color when not in use.[4]
Answer
Change the absorbent now. Time and texture both say it is spent, and the unchanged color does not override them.[1],[4]
Common mistakes
XPressing the oxygen flush with a patient on a non-rebreathing circuit: it sends high pressure gas straight into the airway, so it should never be used.[1],[2]
XLeaving the pop-off valve closed after a leak test or a manual breath: always open it, because a closed valve lets airway pressure climb quickly.[1],[4]
XTrusting the soda lime color alone: the indicator can revert to its original color, so also go by hours of use and granule texture.[1],[4]
XReading the flowmeter at the bottom of the bobbin: read the top of a bobbin or the middle of a ball.[5]
XThinking a non-rebreathing circuit has a CO2 absorbent: it has none and relies on fresh gas flow to clear CO2.[4]
XSkipping the leak test because the machine passed earlier in the day: test before each patient and after changing any circuit part.[1]