Anesthesia oxygen flow and reservoir bag calculator
Work out the oxygen flow for a dog or cat on a rebreathing or non-rebreathing circuit, and the reservoir bag size, from the AAHA and AAFP anesthesia guidelines. A second mode gives the flow for nasal oxygen.
Example
0.5-0.8L/min
20 kg dog. Rebreathing, maintenance. Reservoir bag: 2 L bag, at least 1,200 mL (60 mL/kg), rounded up.
How the example works
- 20 kg × 20-40 mL/kg/min = 400-800 mL/minAAHA: 20-40 mL/kg/min on a rebreathing circuit during maintenance
- 500-800 mL/min ÷ 1,000 mL/L = 0.5-0.8 L/minAAHA sets a minimum of 500 mL/min on a rebreathing circuit, to ensure accurate vaporizer output, so the low end is raised to 500 mL/min
- 20 kg × 60 mL/kg = 1,200 mLReservoir bag: at least 60 mL/kg (5 to 6 tidal volumes), rounded up to the next bag size
For study and for double-checking your own math. The anesthetic and oxygen plan comes from the veterinarian. Watch the patient and the capnograph, and adjust the flow as needed.
Worked example
A 25 kg dog is maintained on inhalant anesthesia through a rebreathing (circle) circuit.
- 25 kg × 20-40 mL/kg/min = 500-1,000 mL/minAAHA: 20-40 mL/kg/min on a rebreathing circuit during maintenance
- 500-1,000 mL/min ÷ 1,000 mL/L = 0.5-1 L/min
- 25 kg × 60 mL/kg = 1,500 mLReservoir bag: at least 60 mL/kg (5 to 6 tidal volumes), rounded up to the next bag size
Set the flowmeter between 0.5 and 1 L/min for maintenance, and use a 2 L reservoir bag: 1,500 mL rounds up to the next bag size.
How it’s calculated
Oxygen flow = weight (kg) × the flow rate for the circuit (mL/kg/min), then ÷ 1,000 for L/min, since flowmeters read in L/min.
| Use | Flow | Source |
|---|---|---|
| Rebreathing (circle), maintenance | 20-40 mL/kg/min, at least 500 mL/min | AAHA |
| Rebreathing, fast depth changes | 2-3 L/min, such as at induction and when the inhalant is turned off (cats: the first 15 minutes) | AAHA, AAFP |
| Non-rebreathing | 200-400 mL/kg/min (cats: at least 200 mL/kg/min, never under 0.5 L/min) | AAHA, AAFP |
| Nasal oxygen (not anesthesia) | 50-100 mL/kg/min, humidified | Merck |
On a rebreathing circuit, AAHA sets a minimum of 500 mL/min so the vaporizer output stays accurate. For patients under 12.5 kg, all of 20-40 mL/kg/min falls below that, so the flow is simply 0.5 L/min. Lower flows waste less oxygen and inhalant and keep in heat and moisture, but anesthetic depth changes more slowly after the vaporizer is turned.
On a non-rebreathing circuit, the high flow is what stops the patient rebreathing exhaled CO2, which is why it’s ten times the rebreathing rate. Some references give 250-500 mL/kg/min; this calculator uses AAHA’s 200-400.
Reservoir bag. For dogs, the bag should hold at least 60 mL/kg, which is 5 to 6 tidal volumes of 10 to 15 mL/kg, rounded up to the next size. Bags of 0.5, 1, 2, 3 and 5 L are common for small animals; a 20 kg dog typically gets a 2 L bag. Cats follow the AAFP table: 0.5 L up to 5 kg and 1 L from 6 to 10 kg on a non-rebreathing circuit, and 1 L for cats of 3 to 10 kg on a rebreathing circuit. Between rows, the calculator uses the next heavier row. Too small a bag collapses when the patient breathes in; too big a bag makes breathing harder to watch.
- SourceGrubb T et al. 2020 AAHA Anesthesia and Monitoring Guidelines for Dogs and Cats. JAAHA 2020;56(2)
- SourceRobertson SA et al. 2018 AAFP Feline Anesthesia Guidelines. J Feline Med Surg 2018;20(7):602-634 (Table 4)
- SourceCooley K. The ins and outs of the anesthesia machine (Proceedings). dvm360, 2009
- SourceLee L. Anesthetic equipment: breathing circuits and scavenging system. Western University of Health Sciences
- SourceQueen’s University Animal Care Committee. SOP 14.5, Anesthetic machine setup, 2022
Choosing the circuit
- AAHA: rebreathing circuits are typically used for patients over 3 to 5 kg. Pediatric rebreathing hoses can be used for smaller patients if a non-rebreathing circuit isn’t available.
- AAFP: cats under 3 kg go on a non-rebreathing circuit; larger cats can use either.
- Never use the oxygen flush valve while a patient is on a non-rebreathing circuit. It sends high-pressure flow straight into the airway.
- Whatever the circuit, check the flow with the capnograph: keep inspired CO2 under 5 mm Hg.
Nasal oxygen
Outside anesthesia, Merck gives 50-100 mL/kg/min of humidified oxygen through a nasal catheter, which delivers 40-60% inspired oxygen while the patient can still be examined and treated. High-flow nasal oxygen, used at specialty centers, is a different system. Sources differ on the most one nasal cannula should carry, so follow the cannula’s instructions and your clinic’s protocol.
Common mistakes
- Leaving the answer in mL/min. 500 mL/min is 0.5 L/min on the flowmeter, not 5.
- Using the rebreathing rate on a non-rebreathing circuit. At 20-40 mL/kg/min, exhaled CO2 builds up in the circuit.
- Forgetting the 500 mL/min minimum for small patients on a rebreathing circuit.
- Rounding the bag down. The bag should hold at least 60 mL/kg, so always round up.
Practice this
Oxygen flow and reservoir bag problems. Type your answer to see the worked solution.
Oxygen flow
A 7-kg dog is maintained on a rebreathing (circle) circuit. AAHA gives 20-40 mL/kg/min, with a minimum of 500 mL/min.
What is the high end of the oxygen flow range, in L/min? Round to 2 decimal places.
Practice numbers, not a dosing reference.
More topics: vet tech math practice.
Practice for the VTNE
Preparing anesthetic equipment is a listed task in the VTNE’s anesthesia domain, and oxygen therapy is one in emergency and critical care.